Hospital X-Ray Machine This
particular hospital X-ray machine is a renovated form available in the series of 300/125
fr. Refurbished hospital X-ray machine is abounded with floor and ceiling rails
comprising a corrected travel of straight arm for tube escalating. In our
manufactured refurbished hospital X-ray machine, the specialty lies in column which has
a complete 360° rotation travel about its vertical axis. Two strong wire ropes
connect the actions of straight arm. Regular accessories like light beam coli
meter is provided along with it for centering. Following
are the certain renovative descriptions:
- Tube used: Bel-THX (DRA1) or
Imported
- Generator (C- Leela): 300 /
125 FR, 300 mA, 125 kV. It is a full wave Rectified X- Ray machine
- Generator C- Leela: 300 / 125
being Full Wave Rectified deliverers the excellent radiographic output.
Large
Focus:
- 300mA-from
40 to 90 KV for 0.02 secs. to 0.4 secs (120 mA)
- 200
mA-from 40 to 100 KV for 0.02 secs. to 1.3 secs (260 mA)
- 100
mA from 40 to 125 KV for 0.02 secs. to 3.0 secs (300 mAs)
Small
Focus:
- 150 mA from 40 to 90 KV for
0.02 secs. to 2.0 secs (300 mAs)
- 100 mA from 40 to 100 KV for
0.02 secs. to 5.0 secs (500 mAs)
- 50 mA from 40 to 125 KV for
0.02 secs. to 5.0 secs (250 mAs)
Its
controlling features are as follows:
- Mains on-off switch
- Line voltage compensation control
- mA
selector switch at 50 mA, 100mA (Small Focus)
- 100
mA, 200 mA, 300 mA (Large Focus)
- Major-40 KV, 50 KV, 60 KV, 70
KV, 80 KV, 90 KV, 100 KV, 115 K
- Minor-2 KV, 4 KV, 6 KV, 8 KV,
10 KV
- 23 Steps electronic timer
from 0.04 to 5.0 secs.
- Bucky on-off switch
- X- Ray and over load
indication lamps
- Double step exposure switch
|  |
 |
|
|