Construction : Double wall construction, interior fabricated of stainless steel(S.S.304). Exterior of corrosion resistant galvanized sheet, finished in powder coated epoxy paint. Full size glass inner door to inspect samples without affecting the chamber temp. Outer double walled metal door with magnetic gasket and lock. Unit mounted on castor wheels for easy movement.
Insulation : 70mm polyurethene insulation (PUF) to ensure better insulation and less leakage of temperature or RH from inner body to the surroundings.
Air Circulation : Temperature is maintained by a quiet running blower circulation air through out the chamber. Forced air circulated vertically down and re circulated through out the chamber for uniform temperature and humidity.
Humidity : Humidity created by steam injection method. The boiler tank is fitted in the back side of the chamber for better servicing, Electro magnetic switch for controlling the wet heater from burning off if water level is not adequate. Float valve provided to control water level in the boiler tank.
Heating & Cooling System : Long lasting SS tubular heaters used as heating element. The stainless steel fins provided to ensure better heat transfer. Hermetically sealed compressor CFC free compressor (134 A gas) coupled with evaporation coil and condensor.
Trays : Removable perforated trays provided made of S.S. wire mesh
Control : Microprocessor based temp and direct RH digital PID controller. Humidity directly measured in % RH by electronic sensor.
Safety Features : High temperature safety cut off, Low water level boiler cut off. Electrical overload cut off. Time delay for compressor switch ON. Electrical circuit breaker
| Temp. Range | 10 to 70°C, Accuracy +/- 0.5°C |
| Humidity Range | 40% RH to 95% RH Accuracy +/- 3% RH |
| Chamber size | Cu. | Trays |
| H W D | Ft. | |
| 45 X 45 X 45 | 3 | 2 |
| 70 X 50 X 50 | 6 | 3 |
| 90 X 60 X 60 | 12 | 3 |
| 125X 60 X 60 | 16 | 4 |
Conditions as per ICH Guidelines. - 25 Deg.C - 60% RH
- 40 Deg.C - 75% RH
- 30 Deg.C - 65% RH
OPTIONAL FITTINGS AT EXTRA COST A. i) Digital Controller with Printer Interface to connect EPSON dot matrix line printer. Print Interval programmable, can print date, time, temp. and humidity.
ii) -Do- but with standard PC software and connectivity.
21 CFR Part 11 Compliance Software
B. Datalogger : 4 point temp. + 4 point humidity datalogger with sensors placed at specific points in chamber with printer interface and memory of 5000 readings per channel.
21 CFR Part 11 Compliance Software
C. Additional high temperature safety system to cut-off all heaters with audio-visual alarm ensuring temp. within chamber doesn't exceed +70°C.
D. Voltage Stabilizer for above unit.
E. EPSON dot matrix 80 column Printer with stand.
F. Standby Humidity and Refrigeration system.
G. 10 Ltr. capacity aspirator bottle(PP) for storage of Mineral water and to serve as input water source for the Water tank/ Steam boiler of Oven.
H. Silicon rubber tubing - 3 meter for water supply from reservoir to Boiler.