Total hydrocabon analyzer for N2, Ar, He, and Air
General characteristics
- Continuous measurement system based on a flame ionisation detector.
- System controlled by microprocessor.
- Easy to use.
- Easy maintenance.
- Weak sample flow.
- Less power consumption.
- CE Analyser.
- Alphanumeric screen 8 x 40 characters, 64 x 240 pixels
- Multifunctional and real-time software.
Dimensions
Height .....4U
Width.....19'
Depth.... 450 mm
Technical characteristics
Three ranges:
0-1 ppm...display resolution 0.01 ppm
0-10 ppm ... display resolution 0.1 ppm
0-100 ppm ... display resolution 0.1 ppm
Accuracy : 0.5% of the reading scale
Detection limit < 0.05 ppm on the range 0-1 ppm
Drift:... 1 % during 24 hours
(0.5 % over a longer period)
Gas connections
Sample.... 1 /8" Swagelok SS
Air ... 1/8" Swagelok SS
H2 ... 1/8" Swagelok SS
Sample flow ... ± 3 l/h
Max Input Pressure... working pressure from 0.3 to 15 Bar
Synthetic air
Flow... 20 l/h
Pressure... 2 Bar
Quality... 5.0
• Hydrogen
Flow...2 l/h
Pressure..... 1 Bar
Quality...6.0
Working Temperature ... ambient and stable temperature (± 20°C)
Power supply...220 Vac, 50-60 Hz
Consumption...500 VA
Output 4-20 mA... 1 output for THC impurity
Connector RJ-45 ... computing maintenance of the system
Output relays ... (dry contacts, SPST 2A / 250 Vac)
1 STATUS contact relay (security alarm)
3 RANGE contact relays (active range)
1 process alarm contact relay (level 1)
1 process alarm contact relay (level 2)
2 contacts for automatic calibration