Calculiclast is intra corporeal contact type lithotripsy system that is effectively used for urinary stone fragmentation of any size. Designed using latest technology, these are acknowledged for their compact designs and high efficiency. Light in weight, our range is virtually maintenance-free and can be availed in varied specifications. Features - Calculiclast is the most effective equipment for urinary stone fragmentation of any size.
- It can fragment renal, bladder or ureter stones with the help of different probes.
- It has adjustable power of the blast.
- Four different preset frequencies for different applications.
- Its hand tool is light-weight & easy to assemble.
- The complete unit is portable & comes in specially designed PVC molded bags of the size 43x32x15 cm
- It is virtually maintenance-free
Components One set of calculiclast consists of: - Control unit
- Two hand pieces
- Silicon rubber tubing: Length 2 meter to attach the hand piece to the control unit.
- Foot switch with 3 meter cable
- Oxygen gas regulator
- PU-6 tubing of length 2.5 Meter to attach the regulator to control unit
- Probes-0.8 mm, 1 mm, 1.2 mm each for ureteric stones, 1.5 mm, 2 mm each for bladder & renal stones
- One HDP molded bag with foam lining inside to accommodate all the parts of calculiclast
Accessories - Adapter to connect hand piece to storz, wolf or olympus make ureternoscopes.
- Suction device for PCNL & URS.
- Modified cystoscope bridge.
Specifications - Control Unit
- Supply voltage 230 Volts A.C. at 50 Hertz
- Input pressure Maximum 6 kg. per cm square
- Pressure source Pressurized Oxygen or filtered pressurized air
- Dimensions Length-220 mm, width-160 mm & height-110 mm, weight- 2.3 kg
- Hand piece: Length - 18.5 cm, Weight-150 gm
- Probes: 0.8mm, 1mm, 1.2mm drawn from 2mm SS wire(length-695mm, 610mm, 610mm respectively), 1.5mm drawn from 2mm SS wire(length-450mm), 2mm(length-410mm)
|